Print

Study on Pulsed X-ray Secondary Standard Ionization Chamber and Calibration Technology

论文摘要

<正>In order to measure the dose of pulsed reference field,the theoretical calculation,Monte Carlo simulation and experiment methods were used to develop a secondary standard ionization chamber for millisecond pulsed X-ray.A pulsed X-ray reference radiation field was built based on a millisecond pulsed X-ray machine,and pulsed response characteristics test of active ionizing radiation dosimeter were carried out.

论文目录

文章来源

类型: 期刊论文

作者: GAO Fei,XU Yang,NI Ning

来源: Annual Report of China Institute of Atomic Energy 2019年00期

年度: 2019

分类: 工程科技Ⅱ辑

专业: 核科学技术

单位: Radiation Metrology Division

分类号: TL811.1

页码: 228

总页数: 1

文件大小: 65K

下载量: 1

相关论文文献

本文来源: https://www.lunwen90.cn/article/4eb2a385add99e15b5105eff.html